The Thrust Plate Prosthesis : Long-term clinical and radiological results


Published online: Jun 27 2013

Leon W. DIEDERIX, Peter J.T.S. VAN WINTERSWIJK, Sander B. SCHOUTEN, Pieter A.G.M. BAKX, Jaap HUIJ

From the Maasstad Hospital, Rotterdam, The Netherlands

Abstract

The Thrust Plate Prosthesis is a femoral implant designed for total hip arthroplasty, based on the principles of physiologic loading of the metaphysis of the proximal femur, and preserving the bone stock. This study presents the long-term clinical and radiological results of 34 patients with 36 Thrust Plate Prostheses. In a retrospective analysis, we investigated the reoperation-free survival as well as the clinical and radiological results. Mean age at operation was 51 ± 6.4 years. Mean follow-up length was 11.9 ± 1.6 years. Reoperation-free survival was 88.9%. Four (11.1%) reoperations were performed, in three patients due to aseptic loosening and in one patient because of a fracture distal to the lateral plate. Three of the reoperations were performed between 12 and 32 months postoperatively. The major complaint was pain at the lateral side of the hip (44%). Radiolucencies did not exceed 1 mm, but 35% of the hips showed resorption of the cortex directly under the thrust plate, together with cancellous bone hypertrophy at the calcar, noted in 97%. Because of the relatively high reoperation-free survival and favourable radiological results, the Thrust Plate Prosthesis appears as a possible alternative to stemmed total hip arthroplasty, especially in relatively young patients.