Reliability of patient specific instrumentation in total knee arthroplasty
Published online: Dec 27 2015
Harold JENNART, Marie-Ange NGO YAMBEN, Theofylaktos KYRIAKIDIS, David ZORMAN
Department of Orthopaedic and Traumatology, CHU Tivoli, Belgium
Abstract
The aim of this study was to compare the precision between Patient Specific Instrumentation (PSI) and Conventional Instrumentation (CI) as determined intra-operatively by a pinless navigation system. Eighty patients were included in this prospective comparative study and they were divided into two homogeneous groups. We defined an original score from 6 to 30 points to evaluate the accuracy of the position of the cutting guides. This score is based on 6 objective criteria. The analysis indicated that PSI was not superior to conventional instrumentation in the overall score (p = 0.949). Moreover, no statistically significant difference was observed for any individual criteria of our score. Level of evidence II.