How close does an uncemented hip stem match the final rasp position ?
Published online: Dec 27 2004
Marco BARINK, Hjalmar MEIJERS, Maarten SPRUIT, Christoph FANKHAUSER, Nico VERDONSCHOT
From the University Medical Centre Nijmegen, The Netherlands and Mathys Ltd Bettlach, Switzerland
Abstract
During total hip arthroplasty the final clinical position
of the cementless CLS stem (Centerpulse) is not
always identical to the position of the final rasp with
which a successful trial reduction was performed.
Therefore, the rasp-stem correspondence of CLS system
(Centerpulse) was investigated in a laboratory
study and compared to the CBC-T system (Mathys).
Both systems showed an average rasp-stem mismatch
below 2 mm in three orthogonal directions. It
was found that this mismatch related to geometric
differences at the corners between rasp and stem.
The measured mismatch is not expected to have
adverse clinical consequences.